The reported events of lead dislodgement, lead sensing problem, and fracture were not confirmed.As received, a complete lead was returned in one piece with the lead helix retracted and clogged with blood.After cleaning and applying torque directly to the connector pin, the helix was able to be extended and retracted.The measured full helix extension length was within product specification.Visual and x-ray examination did not find any anomalies.Electrical testing did not find any indication of conductor fractures or internal shorts.No anomalies were found.
|