• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number LDA210Q/58
Device Problems Fracture (1260); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2024
Event Type  Injury  
Event Description
It was reported that the patient presented remotely via merlin.Net.Upon review, it was found that patient's right ventricular (rv) lead exhibited r wave amplitude variation.It was further noted from x-ray that the lead was dislodged.No intervention was done.There were no patient consequences.
 
Event Description
New information received notes that the lead was explanted and replaced.The patient was stable.
 
Manufacturer Narrative
The reported events of lead dislodgement, lead sensing problem, and fracture were not confirmed.As received, a complete lead was returned in one piece with the lead helix retracted and clogged with blood.After cleaning and applying torque directly to the connector pin, the helix was able to be extended and retracted.The measured full helix extension length was within product specification.Visual and x-ray examination did not find any anomalies.Electrical testing did not find any indication of conductor fractures or internal shorts.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18792750
MDR Text Key336372307
Report Number2017865-2024-33592
Device Sequence Number1
Product Code NVY
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model NumberLDA210Q/58
Device Lot NumberA000149314
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/12/2024
Initial Date FDA Received02/27/2024
Supplement Dates Manufacturer Received03/14/2024
04/16/2024
Supplement Dates FDA Received04/01/2024
04/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GALLANT
Patient Outcome(s) Required Intervention;
Patient Age49 YR
Patient SexMale
-
-