Model Number 6947M55 |
Device Problems
Over-Sensing (1438); Pacing Problem (1439)
|
Patient Problem
Dizziness (2194)
|
Event Date 02/07/2024 |
Event Type
Injury
|
Event Description
|
It was reported that during use of the right ventricular (rv) lead the patient reported light headedness. the rv lead appeared to b e oversensing during the atrial arrhythmia, caused under pacing in ventricular and low r waves noted. a rv lead integrity alert (lia) triggered for two or more high rate non-sustained episodes. the rv lead remains in use. no further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Continuation of d10: dtma1qq crt-d implanted: (b)(6) 2024 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that approximately eight days later the patient had a device check.The patient light headedness was related to the reported rv lead pacing and sensing issues; the rv lead was oversensing atrial fibrillation (af), causing pacing inhibition in a pacemaker dependent patient.Rv lead revision was performed.
|
|
Search Alerts/Recalls
|
|