VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY
|
Back to Search Results |
|
Model Number 3100A VENT,115V,60HZ,DOM |
Device Problem
Inaccurate Flow Rate (1249)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/05/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
H10: a vyaire fsr (field service representative) evaluated the device onsite and troubleshoot the problem, and found the i-time panel meter to be intermittent.Fsr replaced the i-time panel meter and performed a 2k pm.This unit passed the circuit calibration and performance verification.Fsr ran the ventilator for 1 hour and could not duplicate the problem.Ventilator now is ready for patient use.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
It was reported to vyaire medical that the it (inspiratory time) is fluctuating on this unit.It was on a patient when they noticed it was fluctuating.Furthermore, there was no patient harm reported associated with the event.Alternative ventilation was provided as per the customer.
|
|