• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065752914
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Corneal Edema (1791); Visual Impairment (2138)
Event Type  Injury  
Event Description
A customer reported that "90-95%" of the patients experienced the events corneal edema and unexpected visual outcome after cataract surgery.An operating cataract console and handpiece were used to perform the procedures.Patients with corneal edema have not fully recovered.Furthermore, additional information was received from the reporter informing that the patients were treated with corticosteroids and sodium chloride eye drops.Additional information was requested from the reporter regarding patients' status.Follow-up information was received on the patients involved in this event; subsequent mdrs will be filed for each identified patient.
 
Manufacturer Narrative
It was reported that these events occurred with three centurion active sentry handpieces, however it is not known by the initial reporter which handpiece contributed to each event.The sample has been received and in-house evaluation is in progress.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is:(b)(4).
 
Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.A handpiece (hp) was returned for testing on this investigation.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The hp was received for testing on this investigation.A visual assessment of the returned sample revealed a dented nosecone and a worn red connector dot.The handpiece was connected to a calibrated resistance breakout test box, where the connection between the input and output impedance of the sensor was found to be within specifications.A flow rate test was performed on the irrigation and aspiration lines of the handpiece which found the handpiece to meet specifications.The returned sample was connected to a calibrated system.The handpiece tuned successfully and completed a five-minute burn-in test with the system set at 100% ultrasonic and torsional power.The handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements which found the handpiece to meet specifications.The handpiece was found to functionally exhibit no issues during sample testing.Therefore, the root cause of the reported events remains inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18797474
MDR Text Key336448906
Report Number2028159-2024-00314
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065752914
Device Lot Number15PJH6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received02/28/2024
Supplement Dates Manufacturer Received03/04/2024
Supplement Dates FDA Received04/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM (ACTIVE SENTRY)
Patient Outcome(s) Required Intervention; Other;
-
-