ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065753041
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Eye Injury (1845)
Event Type  malfunction  
Event Description
A healthcare professional reported that an ophthalmic operating console had fluid/air exchange problem during surgery.The patient experienced eye injury (unspecified).The type of procedure was not reported.Additional information was requested, none received till date.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The company representative was able to replicate the reported event.The pneumatics module was replaced to address the fluid air exchange (fax) issue.The system was tested and found to meet product specifications.Based on the information obtained, the root cause of the reported eye injury is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The root cause of the reported ¿fluid/air exchange event¿ is attributed to the nonconforming pneumatic module.However, based on the information obtained, the root cause of the reported eye injury is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18797649
MDR Text Key336510251
Report Number2028159-2024-00315
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065753041
Device Lot Number149CV8
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/26/2023
Initial Date FDA Received02/28/2024
Supplement Dates Manufacturer Received03/12/2024
Supplement Dates FDA Received04/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-