MEDTRONIC EUROPE SARL AZURE XT DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number W2DR01 |
Device Problems
Premature Discharge of Battery (1057); Pocket Stimulation (1463)
|
Patient Problem
Device Overstimulation of Tissue (1991)
|
Event Date 10/25/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the right atrial (ra) lead exhibited low impedance.It was also reported that the implantable pulse generator (ipg) exhibited an inaccurate longevity estimate due to a lead impedance measurement.It was reported that the programmer displayed an inaccurate longevity battery estimate for the implantable pulse generator (ipg).The status of the programmer is unknown.It was further reported that the programmer remains in use.It was further reported that the patient experienced pocket stimulation when programmed to unipolar atrial pacing, so the ra lead was kept in bipolar configuration.The ipg exhibited premature battery depletion due to the low ra lead impedance.The ra lead and ipg remain in use. no further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Continuation of d10: model: 4568-53 lead; implanted: (b)(6) 2021 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated that the battery had not yet reached its recommended replacement time.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|