FARAPULSE, INC. FARAWAVE PULSED FIELD ABLATION CATHETER, 35MM; CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER
|
Back to Search Results |
|
Lot Number 8011304-06 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Upon device return and inspection, no outer abnormalities were observed.The device was inspected on the x-ray to look for any abnormalities that could cause deployment issues.Some kinking of the lead wires was observed.The catheter was returned with a guidewire still inserted.An attempt was made to deploy the device.It was not possible to reach full flower deployment.The device was dissected to look for any abnormalities that could have contributed to the unsuccessful deployment.Adhesive residue was found on the flush lumen, causing the lumen to be stuck to the inside of the catheter shaft (image #4).The flush lumen was also broken and kinked, likely as a result of being adhered to the shaft.The reported event was confirmed.
|
|
Event Description
|
Reportable based on analysis completed on 13-feb-2024.Clinical patient id: (b)(6)./ clinical study: advantage af phase ii pf106.It was reported that a farawave ablation catheter 35mm ide during procedure to treat a persistent atrial flutter, after isolating 2 lateral veins, the physician was unable to deploy the device from basket to full flower position, just a biscuit shape was achieved.Prior to that point the doctor was very quick and may have inadvertently tried to deploy the catheter with the marker in the sheath.3 attempts were made to return catheter to neutral then redeploy but were unsuccessful.To solve the issue the procedure was completed without patient complications using another 35mm farawave.The catheter is expected to be returned.However, analysis of the returned device reveled a broken flush lumen.
|
|
Search Alerts/Recalls
|
|
|