• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS SPECTRA CONCEALABLE PENILE PROSTHESIS; PROSTHESIS PENILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS SPECTRA CONCEALABLE PENILE PROSTHESIS; PROSTHESIS PENILE Back to Search Results
Model Number UNK-P-SPP
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  Injury  
Manufacturer Narrative
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available, the cause that contributed to the reported event cannot be established; a conclusion code of cause not established was assigned to this investigation.
 
Event Description
It was reported that one spectra malleable penile prosthesis broke.Therefore, the patient underwent surgery to replace the device.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS SPECTRA CONCEALABLE PENILE PROSTHESIS
Type of Device
PROSTHESIS PENILE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18803611
MDR Text Key336505992
Report Number2124215-2024-11875
Device Sequence Number1
Product Code FAE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUNK-P-SPP
Device Catalogue NumberUNK-P-SPP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received02/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age75 YR
Patient SexMale
-
-