Model Number OER-PRO |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/13/2024 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus the endoscope preprocessor disinfectant solution counter does not reset.The issue was found during preparation for use before a procedure.There was no report of patient injury or medical intervention associated with this event.
|
|
Manufacturer Narrative
|
The device will not be returned to olympus for evaluation.When the customer contacted the service business center to report the event, they were informed the device was at end of life.They were provided with self-help resources to troubleshoot the issue.A supplemental report will be submitted if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The equipment used at the facility is more than 12 years old.It is likely that the board deteriorated over time, causing the disinfectant counter to malfunction accidentally.Olympus will continue to monitor field performance for this device.
|
|