MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
|
Back to Search Results |
|
Model Number 990063-020 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/28/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id: afapro28; product type: balloon catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, a system notice was received indicating that the safety system detected a compromised outer vacuum.The balloon catheter was replaced which resolved the issue.Additionaly, the mapping catheter was reportedly kinked.The mapping catheter was replaced to resolve the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the 990063-020 mapping catheter with lot number 227669010 was returned and analyzed.Visual inspection of the loop segment area showed the loop was intact with no apparent issues.No damage was observed along with the tip/loop section of the mapping catheter.Visual inspection of the pebax tubing area showed the pebax tubing was intact with no apparent issues.No damage was observed along with the pebax tubing section of the mapping catheter.Visual inspection of the electrode(s) showed the electrode(s) were intact with no apparent issues.All electrodes exist on the loop section and no cosmetic issue or anomalies were identified.Visual inspection of the shaft segment area showed that the shaft was kinked approximately 52.5 inches from the lemo connector.Visual inspection of the introducer showed the introducer was intact with no apparent issues.No damage or any other issue was observed along with the introducer.Visual inspection of the lemo connector showed the lemo connector was intact with no apparent issues.No damage or any other issue was observed along with the lemo connector.The insertion compatibility test was performed but unable to be carried out due to a shaft kink.In conclusion, the mapping catheter failed the return product inspection due to a kink/twist was observed on the catheter shaft.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|