Model Number BREVDRV-02 |
Device Problem
Failure to Obtain Sample (2533)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/13/2024 |
Event Type
Injury
|
Event Description
|
It was reported that during a brevera procedure on (b)(6) 2024, a driver error occurred during the procedure, the biopsy could not be completed and the patient was rescheduled for a new procedure.No other information is available.
|
|
Manufacturer Narrative
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.H3: a field engineer examined the equipment and confirmed the operational damage to the driver.The field engineer replaced the biopsy driver and performed a start-up driver test.The system passed all tests and meets manufacturer specifications.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.
|
|