It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4, a thickened anterior leaflet, and previous cardiac surgery.An ntw clip (30325r1106) was inserted, however, it was unable to grasp the anterior leaflet.The anterior gripper had a gripper actuation issue.Troubleshooting was performed, but the issue was unable to be resolved.Therefore, the clip was removed and replaced.A second ntw clip (31026r1007) was inserted, however, the gradient increased to 13mmhg.Therefore, the clip was removed, and the procedure was discontinued.Mr remained at a grade of 4 and the gradient returned to baseline.There were no adverse patient effects and no clinically significant delay in the procedure.
|
All available information was investigated and the reported gripper actuation issue was not confirmed via returned device analysis.The reported positioning failure of inability to grasp could not be replicated in a testing environment as they were related to patient and/or procedural conditions.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation determined the reported actuation issue and inability to grasp the leaflets appear to be related to patient morphology/pathology due to the thick anterior leaflet.There is no indication of a product issue with respect to manufacture, design or labeling.
|