• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-NTW
Device Problems Positioning Failure (1158); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.Na.
 
Event Description
It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4, a thickened anterior leaflet, and previous cardiac surgery.An ntw clip (30325r1106) was inserted, however, it was unable to grasp the anterior leaflet.The anterior gripper had a gripper actuation issue.Troubleshooting was performed, but the issue was unable to be resolved.Therefore, the clip was removed and replaced.A second ntw clip (31026r1007) was inserted, however, the gradient increased to 13mmhg.Therefore, the clip was removed, and the procedure was discontinued.Mr remained at a grade of 4 and the gradient returned to baseline.There were no adverse patient effects and no clinically significant delay in the procedure.
 
Manufacturer Narrative
All available information was investigated and the reported gripper actuation issue was not confirmed via returned device analysis.The reported positioning failure of inability to grasp could not be replicated in a testing environment as they were related to patient and/or procedural conditions.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation determined the reported actuation issue and inability to grasp the leaflets appear to be related to patient morphology/pathology due to the thick anterior leaflet.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18807915
MDR Text Key337192008
Report Number2135147-2024-00932
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/25/2024
Device Catalogue NumberCDS0702-NTW
Device Lot Number30325R1106
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received02/29/2024
Supplement Dates Manufacturer Received03/04/2024
Supplement Dates FDA Received03/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
-
-