Model Number 1944/52
Device Problems Break (1069); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Event Description
During follow-up, noise leading to oversensing was observed on the right ventricular (rv) lead and right atrial (ra) lead.The oversensing on the ra lead led to inappropriate automatic mode switch.Lead damage was suspected but was not confirmed visually.Noise could be reproduced through provocative testing.It was suspected that the oversensing was due to myopotentials.The device was reprogrammed to resolve the event.The patient was stable and will continue to be monitored.There were no adverse consequences.Related manufacturer reference number: 2017865-2024-33795.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
Permanent defibrillator electrodes
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18810739
MDR Text Key336615150
Report Number2017865-2024-33792
Device Sequence Number1
Product Code NVY
Combination Product (y/n)Y
Reporter Country CodePO
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Model Number1944/52
Device Lot NumberP000069068
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/14/2024
Initial Date FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-