BIOTRONIK SE & CO. KG SOLIA S 53; PACING LEAD Back to Search Results
Model Number 377177
Device Problems Failure to Capture (1081); Low impedance (2285); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/05/2024
Event Type  malfunction  
Event Description
Lead had inadequate sensing, no capture, and impedance less than 200 in post op recovery after implant.Patient was returned to ep lab and lead was successfully repositioned.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Manufacturer Narrative
Lead was explanted and replaced due to dislodgement on (b)(6) 2024.The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLIA S 53
Type of Device
PACING LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18811867
MDR Text Key336793899
Report Number1028232-2024-01128
Device Sequence Number1
Product Code NVN
UDI-Device Identifier04035479118266
UDI-Public(01)04035479118266(17)251130
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P950037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number377177
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/20/2024
Initial Date FDA Received02/29/2024
Supplement Dates Manufacturer Received03/04/2024
03/26/2024
Supplement Dates FDA Received03/04/2024
04/01/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age70 YR
Patient SexMale
-
-