Model Number MAJ-2111 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation.A visual inspect of the connecting tube ensured that there were no cracks, breaks, rips, scratches, attached debris, or other irregularities.Only one lever was broken.The broken portion was not returned.In addition, the pin was inspected with no bends or breaks.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the connecting tube separated hook part kept breaking off.The issue was observed during reprocessing.There were no reports of patient harm.Related patient identifier: (b)(6).
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the exact manufacturing date was unknown (based off the lot/serial number, the manufacturing date is on an unknown day in august of 2021).However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely that external stress was applied to the lock lever of connecting tube, which broke the lever and the tube was unable to be connected.As a cause of external stress, the user may have applied force toward incorrect direction.However, the definitive root cause was unable to be determined.The event can be prevented by following the instructions for use (ifu) which state: "preparation and inspection -visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.-move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|