Model Number MAJ-2111
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.A visual inspect of the connecting tube ensured that there were no cracks, breaks, rips, scratches, attached debris, or other irregularities.Only one lever was broken.The broken portion was not returned.In addition, the pin was inspected with no bends or breaks.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the connecting tube separated hook part kept breaking off.The issue was observed during reprocessing.There were no reports of patient harm.Related patient identifier: (b)(6).
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the exact manufacturing date was unknown (based off the lot/serial number, the manufacturing date is on an unknown day in august of 2021).However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely that external stress was applied to the lock lever of connecting tube, which broke the lever and the tube was unable to be connected.As a cause of external stress, the user may have applied force toward incorrect direction.However, the definitive root cause was unable to be determined.The event can be prevented by following the instructions for use (ifu) which state: "preparation and inspection -visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.-move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Type of Device
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18812474
MDR Text Key336644031
Report Number9610595-2024-04295
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170404061
UDI-Public04953170404061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190969
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-2111
Device Lot Number38A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/29/2024
Supplement Dates Manufacturer Received04/08/2024
Supplement Dates FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-