• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. SYMMETRY GEN MODEL 8103; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. SYMMETRY GEN MODEL 8103; GENERATOR Back to Search Results
Model Number 8103
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Depression (2361)
Event Date 02/07/2024
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported that the patient was initially getting efficacy from the device but since titrating the device the patient notes to be getting worse.No other relevant information has been received to date.
 
Manufacturer Narrative
H3.Device evaluated by mfr?; code 81, device evaluation and return of the device is not necessary.It will not add value to the investigation as the root cause is known.
 
Event Description
Physician's response received.The cause of the increased depression is due to depleted battery (normal eos, 0% battery) and the depression level rise was below pre-vns baseline levels.Patient had a battery replacement.Device evaluation is not necessary as it will add no value to the investigation, the root cause is known.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMMETRY GEN MODEL 8103
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key18813124
MDR Text Key336798053
Report Number1644487-2024-00239
Device Sequence Number1
Product Code MUZ
UDI-Device Identifier05425025750504
UDI-Public05425025750504
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/28/2022
Device Model Number8103
Device Lot Number205219
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received02/29/2024
Supplement Dates Manufacturer Received04/15/2024
Supplement Dates FDA Received04/16/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age29 YR
Patient SexFemale
-
-