Model Number 86700J
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Event Description
It was reported that a visions pv.018 catheter was used in a peripheral therapeutic procedure in the distal sfa.While passing through the lesion, the image disappeared.No patient injury reported.During the product return evaluation, missing material from the expanded single lumen was observed.This product problem is being submitted due to the missing material.There is a potential for harm if the malfunction were to recur.
 
Manufacturer Narrative
This complaint was reviewed and investigated according to the manufacturer¿s policy.Blocks a2-a5: no information available.Blocks b6 & b7: no information available.Block c: not applicable for this device.Blocks d6 & d7: not applicable for this device.Blocks e1 & g2: country is japan.Block h6: the probable cause of the missing material is damage during use/handling as evidence by the sanguineous material observed within the lifted and missing expanded single lumen portion.Manipulation, strain, impact, and forces associated with use can affect the integrity of the device.Blocks h7 & h9: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONS PV .018 DIGITAL IVUS IMAGING CATHETER
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
VOLCANO CORPORATION
3721 valley centre drive #500
san diego CA 92130
Manufacturer (Section G)
VOLCARICA S.R.L.
coyol free zone &business park
b37
alajuela
CS  
Manufacturer Contact
ayse kharodawala
3721 valley centre drive #500
san diego, CA 92130
858720-406
MDR Report Key18813294
MDR Text Key336989961
Report Number3008363989-2024-00016
Device Sequence Number1
Product Code OBJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K150442
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number86700J
Device Catalogue Number86700J
Device Lot Number0302802545
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDE WIRE, MFG AND SIZE UNK; INTRODUCER SHEATH MFG AND SIZE UNK
-
-