Model Number RFIT-ASY-0147 |
Device Problem
False Positive Result (1227)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation: a potential false positive c.Tropicalis result on the biofire bcid2 panel was reported after testing a patient blood culture sample.The patient received antifungal treatment due to the biofire bcid2 result.The final diagnosis was fungemia due to c.Tropicalis.No serious injury or death reported.Conclusion: the investigation determined that the most likely cause of the discrepant c.Tropicalis result was the presence of non-viable organism/nucleic acid in the blood culture media bottle.Similar events were recently observed in the field with the combination of the biofire bcid2 panel and bd bactec¿ blood culture bottles.A field safety corrective action (fsca) was issued in the u.S.On (b)(6) 2024, and countries outside u.S.On (b)(6) 2024, via fsca #(b)(4) for this issue.While blood culture vials are autoclaved and routinely quality controlled for sterility, non-viable organisms or nucleic acids can remain in the blood culture media after the sterilization process.The presence of non-viable organisms and nucleic acid does not compromise the intended use of blood culture media, culturing viable microorganisms; however, the biofire bcid2 panel does not distinguish between nucleic acid from viable or non-viable organisms.The biofire bcid2 panel is working as intended.The "laboratory precaution" and "limitation" sections of the biofire bcid2 panel instructions for use (ifu) (www.Online-ifu.Com/iti0048) outlines the potential for false positive detections during molecular testing with sterile blood culture media containing detectable levels of non-viable organisms and/or nucleic acid.Importantly, results from the biofire bcid2 panel are intended to be correlated with the clinical history, epidemiological data, and other data available to the clinician evaluating the patient.All identification results provided by the biofire bcid2 panel are intended to be interpreted in conjunction with gram stain results.Clinical performance: according to table 38.Biofire bcid2 panel clinical performance summary, candida spp.Of the biofire bcid2 panel ifu, the performance claim for the c.Tropicalis assay compared to standard of care identification for genus level followed by pcr & sequencing of isolates for species identification showed an overall sensitivity of (b)(4) ((b)(4)) and an overall specificity of (b)(4) ((b)(4)).The single false positive specimen was identified as a cross-reactivity between the biofire bcid2 panel c.Tropicalis assay and high titer candida parapsilosis.This cross-reactivity is a known limitation of the biofire bcid2 panel.
|
|
Event Description
|
Summary: (b)(6) ((b)(6) colombia) reported a potential false positive candida tropicalis result on the biofire blood culture identification 2 (bcid2) panel after testing a patient blood culture sample.Due to the biofire bcid2 panel result, the patient may have received inappropriate treatment.The investigation determined that the most likely cause of the discrepant c.Tropicalis result was the presence of non-viable organism/nucleic acid in the blood culture media bottle.
|
|
Event Description
|
Summary: (b)(6) ((b)(6) colombia) reported a potential false positive candida tropicalis result on the biofire blood culture identification 2 (bcid2) panel after testing a patient blood culture sample.Due to the biofire bcid2 panel result, the patient may have received inappropriate treatment.The investigation determined that the most likely cause of the discrepant c.Tropicalis result was the presence of non-viable organism/nucleic acid in the blood culture media bottle.Update for follow-up report: when this report (fda report # mdr 3002773840-2024-00262) with an internal investigation # (b)(6) (patient #1) was initially submitted by biofire to the fda on (b)(6) 2024, the customer had only provided patient information and had not provided the run file for the biofire bcid2 panel test.Due to this, biofire submitted a report with only patient information.On (b)(6) 2024, biofire received run file from the customer.Upon review of the run file, it was realized that a report had already been submitted to the fda (fda report # mdr 3002773840-2023-00005) on (b)(6) 2023, with an internal investigation # (b)(6) (patient #2).However, at that time, the customer had not provided any patient information and had only provided the run file.Due to this, biofire had submitted a report with only the biofire bcid2 panel run file information.Therefore, two reports were submitted by biofire for the same case.This follow-up report (fda report # mdr 3002773840-2024-00262 follow-up report) includes both the patient and the biofire bcid2 panel run file information.
|
|
Manufacturer Narrative
|
Investigation: a potential false positive c.Tropicalis result on the biofire bcid2 panel was reported after testing a patient blood culture sample.The patient received antifungal treatment due to the biofire bcid2 result.The final diagnosis was fungemia due to c.Tropicalis.No serious injury or death reported.Update for follow-up report: when this report (fda report # mdr 3002773840-2024-00262) with an internal investigation # (b)(6) (patient #1) was initially submitted by biofire to the fda on (b)(6) 2024, the customer had only provided patient information and had not provided the run file for the biofire bcid2 panel test.Due to this, biofire submitted a report with only patient information.On (b)(6) 2024, biofire received run file from the customer.Upon review of the run file, it was realized that a report had already been submitted to the fda (fda report # mdr 3002773840-2023-00005) on (b)(6) 2023, with an internal investigation # (b)(6) (patient #2).However, at that time, the customer had not provided any patient information and had only provided the run file.Due to this, biofire had submitted a report with only the biofire bcid2 panel run file information.Therefore, two reports were submitted by biofire for the same case.This follow-up report (fda report # mdr 3002773840-2024-00262 follow-up report) includes both the patient and the biofire bcid2 panel run file information.Conclusion: the investigation determined that the most likely cause of the discrepant c.Tropicalis result was the presence of non-viable organism/nucleic acid in the blood culture media bottle.Similar events were recently observed in the field with the combination of the biofire bcid2 panel and bd bactec¿ blood culture bottles.A field safety corrective action (fsca) was issued in the u.S.On (b)(6) 2024, and countries outside u.S.On (b)(6), 2024, via fsca #(b)(4) for this issue.While blood culture vials are autoclaved and routinely quality controlled for sterility, non-viable organisms or nucleic acids can remain in the blood culture media after the sterilization process.The presence of non-viable organisms and nucleic acid does not compromise the intended use of blood culture media, culturing viable microorganisms; however, the biofire bcid2 panel does not distinguish between nucleic acid from viable or non-viable organisms.The biofire bcid2 panel is working as intended.The "laboratory precaution" and "limitation" sections of the biofire bcid2 panel instructions for use (ifu) (www.Online-ifu.Com/iti0048) outlines the potential for false positive detections during molecular testing with sterile blood culture media containing detectable levels of non-viable organisms and/or nucleic acid.Importantly, results from the biofire bcid2 panel are intended to be correlated with the clinical history, epidemiological data, and other data available to the clinician evaluating the patient.All identification results provided by the biofire bcid2 panel are intended to be interpreted in conjunction with gram stain results.Clinical performance: according to table 38.Biofire bcid2 panel clinical performance summary, candida spp.Of the biofire bcid2 panel ifu, the performance claim for the c.Tropicalis assay compared to standard of care identification for genus level followed by pcr & sequencing of isolates for species identification showed an overall sensitivity of (b)(4)).The single false positive specimen was identified as a cross-reactivity between the biofire bcid2 panel c.Tropicalis assay and high titer candida parapsilosis.This cross-reactivity is a known limitation of the biofire bcid2 panel.
|
|