• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO ADAPTER MODULE EPK W/OUT EYEPC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO ADAPTER MODULE EPK W/OUT EYEPC Back to Search Results
Model Number PVK-1070Z
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2024
Event Type  malfunction  
Event Description
The time of event is not during procedure.There was no report of patient harm.Video image failure.
 
Manufacturer Narrative
This device is not distributed in us so that 510k# is blank.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO ADAPTER MODULE EPK W/OUT EYEPC
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale, NJ 07645
2015712318
MDR Report Key18813707
MDR Text Key337734756
Report Number9610877-2024-51728
Device Sequence Number1
Product Code FEA
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPVK-1070Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2024
Initial Date Manufacturer Received 02/16/2024
Initial Date FDA Received03/01/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-