• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAMILTON MEDICAL AG HAMILTON MEDICAL AG; HAMILTON-C6

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAMILTON MEDICAL AG HAMILTON MEDICAL AG; HAMILTON-C6 Back to Search Results
Model Number HAMILTON-C6
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2023
Event Type  malfunction  
Event Description
Connectivity lacks between centricity-ge&hamilton.
 
Manufacturer Narrative
The complaint has been reopened and reviewed according to fda form 483 inspectional observation ems #2, eobs2 from the fda inspection conducted between july 17 to july 21, 2022 at the ems and bonaduz sites.A detailed investigation was performed by an expert from the technical service: since the complaint in question was submitted to hamilton medical ag about one year ago, no attempts will be performed to obtain additional information.No further investigation or correction will be performed except those mentioned above.In future hamilton medical ag will report an event similar to this issue as it will be deemed a reportable event.The allegation in this complaint was confirmed to be a complaint.With this investigation it has been confirmed that the device failed to meet its specifications at the time of the event while the ventilator was prepared for use.The root cause was determined to be driver issue on the side of the hospital management system.A restart of the device solved the issue.There was no patient or user harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMILTON MEDICAL AG
Type of Device
HAMILTON-C6
Manufacturer (Section D)
HAMILTON MEDICAL AG
via crusch 8
bonaduz, 7402
SZ  7402
Manufacturer Contact
andre sonnenberg
via crusch 8
bonaduz, graubunden 7402
SZ   7402
MDR Report Key18814295
MDR Text Key336754021
Report Number3001421318-2023-40938
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K201658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHAMILTON-C6
Device Catalogue Number160021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/27/2023
Initial Date FDA Received03/01/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-