During an implant procedure, a high pacing impedance, loss of sensing, and loss of capture were observed on the left ventricular (lv) lead.The lv lead was explanted and replaced to resolve the event.The patient was stable and will continue to be monitored.
|
The reported events of failure to capture, failure to sense, and high pacing impedance were not confirmed.As received, a complete lead was returned in one piece for analysis.Visual inspection and x-ray examination of the lead did not find any anomalies, except for procedure damage.Electrical testing did not find indication of an internal short.Electrical conductor discontinuities found were due to procedural damage.
|