The device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device number lot 31210113l and no internal action related to the complaint was found during the review.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
It was reported that a patient underwent an atrial fibrillation ablation procedure with a qdot micro¿ catheter and the ngen generator displayed the error message: "heat slope too high" when the catheter was connected and the medical team was coming on ablation.The ablation stopped almost immediately.They checked the irrigation and it appeared that something was clogging the pores on the tip of the catheter.The correct catheter settings had been selected.The catheter was replaced, the issue was resolved, and the procedure was continued.No patient consequences were reported.
|