Catalog Number 382523 |
Device Problem
Retraction Problem (1536)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/03/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1.Address information was not provided, therefore, xx was used as a place holder.H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
|
Event Description
|
It was reported that 5 bd insyte autog bc blu 22ga x 1.0in needles wouldn't retract.The following information was provided by the initial reporter: it was reported by customer that the needle will not retract when the button is pushed.They have a video showing this.Only way in which they were able to do this was by moving the catheter.It seems as though a piece of the button is getting stuck on the catheter luer lock.Additional information received: "their pt rn noticed the defect while inserting ivs, their pt rn also said it has occurred roughly 4-5 times while inserting ivs.4.The sample is no longer available.They saw how many were defective and then those needles were disposed of in their sharps container.".
|
|
Manufacturer Narrative
|
Investigation results: a device history record review was completed by our quality engineer team for provided material number 382523 and lot number 3249704.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Complaints received for this device and reported condition will continue to be tracked and trended.
|
|
Event Description
|
No additional information.
|
|
Search Alerts/Recalls
|