• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. INTROCAN SAFETY®; CATHETER,INTRAVASCULAR,THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. INTROCAN SAFETY®; CATHETER,INTRAVASCULAR,THE Back to Search Results
Catalog Number 4251644-020
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2024
Event Type  malfunction  
Event Description
As reported by the user facility: it was reported that the introcan safety pur 20g, 1.1x32mm-us while being used on a patient were "bendy" during insertion on patient and shattered during use.The emergency department has reported that the 20-gauge introscan safety iv catheter lot number 0061864777 has been kinking in patients during the catheter insertion and occasionally shredding when pulling the needle out of the catheter.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number: (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Fifty-nine (59) unused samples were provided for investigation.The provided 59 samples were visually evaluated, and no product deviation was noted.The defect is not confirmed.Retained units were evaluated and passed the internal testing.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Additionally, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Type of Device
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key18823286
MDR Text Key336750581
Report Number2523676-2024-00237
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier04046964398971
UDI-Public(01)04046964398971
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020785
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4251644-020
Device Lot Number0061864777
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/20/2024
Initial Date FDA Received03/02/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-