Medwatch fields d1-d4, g1, and g4 were updated.The field service engineer checked the analyzer and was not able to determine a root cause.He ran instrument checks which passed and loaded a new reagent pack.The customer performed calibration, qc, and precision testing which passed.The customer repeated the affected patient samples and the results were consistent with the correct values.The investigation was not able to identify a specific root cause.The issue was consistent with a handling or other issue with the suspected reagent pack.
|