SYNTHES GMBH EXTRACTSCR F/TIB+FEM NAILS; EXTRACTOR Back to Search Results
Catalog Number 03.010.000
Device Problem Break (1069)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Event Description
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2024, the patient underwent a removal of the tna (tibial nail advanced) nail implants.A screw thread for removal-use screw was crushed, and connecting the nail was difficult.A hexagonal wrench was planned to be used, but it was noticed that both of the instrument sets concerned originally did not include the hexagonal wrench, so this was not possible.Finally, all the nails of inserted implants were removed.The surgery was completed successfully without any delay.The patient outcome was reported to be stable.The reason for the removal of implants is unknown.This report involves one conical extraction screw for ti femoral and tibial nails.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E1 initial reporter facility name: (b)(6).E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D4: lot.H3, h4, h6: part# 03.010.00.Lot # 9045594.Manufacturing site: werk hägendorf.Supplier: na.Release to warehouse date: 07 aug 2014.Expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the threads of the extractscr f/tib+fem nails were found stripped.The observed condition was consistent as an end of life indicator for the device.No other issues where identified.The broken allegation cannot be confirmed.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was unconfirmed as the observed of extractscr f/tib+fem nails would not have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.G1: manufacturing site name and address.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTSCR F/TIB+FEM NAILS
Type of Device
EXTRACTOR
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf CO 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18825485
MDR Text Key336755617
Report Number8030965-2024-03119
Device Sequence Number1
Product Code HWB
UDI-Device Identifier07611819160361
UDI-Public(01)07611819160361
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.010.000
Device Lot Number9045594
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/04/2024
Supplement Dates Manufacturer Received03/05/2024
04/04/2024
Supplement Dates FDA Received03/20/2024
04/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
EXTRACTSCR F/TIB+FEM NAILS; UNK - NAILS: TIBIAL; UNK - WRENCHES
-
-