• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DELTA CER HEAD 12/14 32MM +5; ARTICULEZE HEAD (12/14 TAPER) : HIP CERAMIC FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DELTA CER HEAD 12/14 32MM +5; ARTICULEZE HEAD (12/14 TAPER) : HIP CERAMIC FEMORAL HEADS Back to Search Results
Catalog Number 136532320
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Joint Dislocation (2374); Physical Asymmetry (4573)
Event Date 02/26/2024
Event Type  Injury  
Event Description
It was reported that on (b)(6) 2024, the patient underwent the surgery via tha with the implant in question.In the surgery, the product in question was dislocated after installation.Therefore, the length of the head was changed urgently to cope with the problem.No further information is available.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H6: component code: appropriate term/code not available (g07002) used to capture no findings available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA CER HEAD 12/14 32MM +5
Type of Device
ARTICULEZE HEAD (12/14 TAPER) : HIP CERAMIC FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18825984
MDR Text Key336776166
Report Number1818910-2024-04916
Device Sequence Number1
Product Code LZO
UDI-Device Identifier10603295033448
UDI-Public10603295033448
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K031803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number136532320
Device Lot Number4266142
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNKNOWN HIP ACETABULAR LINERS
Patient Outcome(s) Required Intervention;
-
-