• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP INC. PRESTIGE ATRA GRASPER DBL-ACT 5MM; HANDHELD PRODUCTS & LIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP INC. PRESTIGE ATRA GRASPER DBL-ACT 5MM; HANDHELD PRODUCTS & LIGATION Back to Search Results
Model Number 8360-10
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  Injury  
Manufacturer Narrative
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event Description
It was reported to aesculap inc.That a prestige atra grasper dbl-act 5mm (part # 8360-10) was used during a procedure on (b)(6) 2024.According to the complainant the grasper jaw fell apart inside patient.Reportedly there was an unspecified delay to the surgery due to the time to remove pieces from inside the patient.The adverse event is filed under aic reference (b)(4).
 
Manufacturer Narrative
Updated information: h6 codes updated manufacturing site evaluation: one (1) used sample was returned with visible evidence of damage.The jaws of the device were broken and did not open or close.An approved project is in place to further address issues relating to the prestige grapser jaws breaking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE ATRA GRASPER DBL-ACT 5MM
Type of Device
HANDHELD PRODUCTS & LIGATION
Manufacturer (Section D)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer (Section G)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
jonathan severino
3773 corporate parkway
center valley, PA 18034
4847197287
MDR Report Key18826567
MDR Text Key336861299
Report Number2916714-2024-00007
Device Sequence Number1
Product Code NWV
UDI-Device Identifier04046955083374
UDI-Public4046955083374
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8360-10
Device Catalogue Number8360-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received03/04/2024
Supplement Dates Manufacturer Received02/07/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-