• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2024
Event Type  malfunction  
Event Description
It was reported, the xenon light source exhibited water leak from the output connector socket.There were no reports of patient harm.
 
Manufacturer Narrative
In speaking with the olympus technical assistance center (tac) the customer reported that the device had a water leak.Tac guided the customer in making sure that the water connection was seated properly, as the customer checked the scope was not seated properly.Once the scope was seated correctly the issue no longer occurred.As the issue was corrected, the device will not be returned to olympus.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation and associated h6 coding.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation and the information provided, it is likely that the reported issue occurred due to the scope not being connected correctly by the user.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18827130
MDR Text Key336789383
Report Number3002808148-2024-02001
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/04/2024
Supplement Dates Manufacturer Received04/12/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-