• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO13.2
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2024
Event Type  Injury  
Event Description
The reporter indicated that the surgeon implanted a 13.2mm vticmo13.2; -10.5/2.0/095 (sphere/cylinder/axis) implantable collamer lens into the patient's right eye (od) on (b)(6)2024.The patient experienced excessive vaulting.On this same date in a separate surgery the lens was exchanged with a same length lens but implanted in a different axis.This resolved the problem.The cause of the event was reported as unknown.
 
Manufacturer Narrative
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
 
Manufacturer Narrative
H3: device evaluation: lens was returned in a microcentrifuge vial, dry.Visual inspection found the lens haptic bent/deformed.Dimensional inspection found the lens within specifications.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker ave
monrovia, CA 91016
MDR Report Key18827239
MDR Text Key336791501
Report Number2023826-2024-00958
Device Sequence Number1
Product Code QCB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberVTICMO13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/31/2024
Initial Date FDA Received03/04/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL-SFC-45, LOT#-UNK.; FOAM TIP PLUNGER MODEL-FTP, LOT#-UNK.; INJECTOR MODEL-MSI-PF, LOT#-UNK.
Patient Outcome(s) Required Intervention;
Patient Age27 YR
Patient SexMale
-
-