Catalog Number CDS0702-XTW |
Device Problems
Unintended Movement (3026); Noise, Audible (3273); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that during device preparation of the clip delivery system (cds) for a mitraclip procedure, when the dc handle was rotated the dc shaft moved irregularly.The dc shaft did not rotate smoothly.The dc shaft was rotated clockwise and jolted forward in the direction it was turned, then rotated counterclockwise and jolted again.Rotating the dc handle felt normal, up until the dc shaft jumped and made an audible click noise.Rotation was not smooth at that point.Troubleshooting was performed as per instructions for use (ifu).Device preparation was repeated, but the issue persisted.The device had no contact with the patient.The procedure was completed with a replacement.
|
|
Manufacturer Narrative
|
All available information was investigated, and the reported unintended movement associated with delivery catheter (dc) shaft positioning, noise and physical resistance/sticking of the dc handle could not be confirmed during the returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no other complaints from the lot.Based on information provided and the results of the returned device analysis (unable to confirm the reported issues), a cause for the reported physical resistance/sticking of the dc handle and unintended movement associated with dc shaft positioning resulting in noise could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
|
|
Search Alerts/Recalls
|