Model Number L111 |
Device Problems
Premature Discharge of Battery (1057); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/14/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that this pacemaker was suspected of exhibiting premature battery depletion.Data analysis was performed, and boston scientific technical services indicated that the power consumption is increasing in a gradual fashion and recommended device replacement due to this anomaly.If a new replacement window is desired, the healthcare professional can send new data to have the battery status re-evaluated.There were no adverse patient effects reported.The device remains in-service.
|
|
Event Description
|
It was reported that this pacemaker was suspected of exhibiting premature battery depletion.Data analysis was performed, and boston scientific technical services indicated that the power consumption is increasing in a gradual fashion and recommended device replacement due to this anomaly.If a new replacement window is desired, the healthcare professional can send new data to have the battery status re-evaluated.There were no adverse patient effects reported and the device remains in-service.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Please refer to b5 describe event or problem for more information regarding the specific circumstances of this event.
|
|