MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ESSENTIO MRI DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ESSENTIO MRI DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L111
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Event Description
It was reported that this pacemaker was suspected of exhibiting premature battery depletion.Data analysis was performed, and boston scientific technical services indicated that the power consumption is increasing in a gradual fashion and recommended device replacement due to this anomaly.If a new replacement window is desired, the healthcare professional can send new data to have the battery status re-evaluated.There were no adverse patient effects reported.The device remains in-service.
 
Event Description
It was reported that this pacemaker was suspected of exhibiting premature battery depletion.Data analysis was performed, and boston scientific technical services indicated that the power consumption is increasing in a gradual fashion and recommended device replacement due to this anomaly.If a new replacement window is desired, the healthcare professional can send new data to have the battery status re-evaluated.There were no adverse patient effects reported and the device remains in-service.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Please refer to b5 describe event or problem for more information regarding the specific circumstances of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO MRI DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18828332
MDR Text Key337025729
Report Number2124215-2024-13048
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/14/2018
Device Model NumberL111
Device Catalogue NumberL111
Device Lot Number236163
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2024
Initial Date FDA Received03/04/2024
Supplement Dates Manufacturer Received03/12/2024
Supplement Dates FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2016
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-