H4: the lot was manufactured between may 10-11, 2023.H10: the device was received for evaluation.A visual inspection was performed, and traces of silicone oil were observed on the interior wall of the housing.Normal manufacturing process requires silicone oil to be applied to the exterior of the elastomeric bladder to prevent potential rupture from housing contact.A small amount of silicone oil from the bladder may have dripped onto the interior wall of the housing during manufacturing; this condition is cosmetic and has no impact on the function or safety of the product.Therefore, silicone oil on the interior wall of the housing is an expected product characteristic and there was no product non-conformance.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|