• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-1100
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed that during the device evaluation that there were foreign objects on the adhesive of the bending section cover of the videoscope.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned, and the evaluation found no reportable malfunctions aside from the allegation reported in b5.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that reprocessing was not conducted properly due to leakage from worn s-cover.Subsequently, the materials were not possibly eliminated.However, a definitive root cause could not be determined.Occurrence of the event can be detected/prevented by handling the device according to the following instructions for use (ifu).Detection methods are written in ifu: gif-1100 operation manual "chapter 3 preparation and inspection" prevention measures are written in ifu: gif-1100 reprocessing manual "chapter 5 reprocessing the endoscope" olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18832553
MDR Text Key336862653
Report Number9610595-2024-04497
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K222584
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberGIF-1100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/04/2024
Supplement Dates Manufacturer Received03/08/2024
Supplement Dates FDA Received03/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-