Reportable based on device analysis completed on 14feb2024.It was reported that the device could not cross the lesion.The 87% stenosed target lesion was located in the moderately tortuous and moderately calcified left circumflex artery.A comet ii was selected for use.During the procedure, it was noted that the device could not cross the lesion.The procedure was completed with another of same device.No complications were reported, and patient was stable post procedure.However, device analysis revealed that the device was cut and only 38cm from the tip was returned.
|