MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
|
Back to Search Results |
|
Model Number 9732719 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/22/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi70000028230, serial/lot #: (b)(6); product id: bi71000117, product id: bi71000611.A manufacturer representative (rep) went to the site to test the imaging system.The connector and solid state relay were replaced and the system performed as intended.Codes: b01, c07, d02 g2) foreign country: japan.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used intra-operatively for an unknown procedure.It was reported that there was about 5 millimeters of misalignment on the caudal side with the imaging system navigation.Medtronic imaging and the surgery were not aborted.No further information available.
|
|
Event Description
|
Additional information was received.The impact on the patient was unknown.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
H2) continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000117, serial/lot #: (b)(4).1, product id: bi71000611, serial/lot #: (b)(6).1 h2-3) the mobile view station (mvs) ethernet cable was returned to the manufacturer for evaluation.After visual/physical examination, the reported issue was not confirmed.The results of the analysis concluded that no failure was found.Codes: b01, c19, d14 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
H3, h6) the product id: bi71000611, lot number: 450719 rev.1, was returned to the manufacturer and analysis did not confirm the reported event of "inaccurate information".The x-ray relay failed bench test.Resistance between point 1 and 22 failed, it shorted.Low resistance of 0.4 ohms was measured and the resistance between point 4 and point +3 was 48.40 kiloohm.It was reported the varistor (used for protecting a circuit from high voltage surges) passed.Previously reported codes b01 and d02 are applicable as well as newly reported code, c02.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|