BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING
|
Back to Search Results |
|
Catalog Number H93811 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
E1: initial reporter facility name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was particulate matter (pm) contamination on the tubing of a repeater pump tube set.The pm was further described as, "a black line was immediately noticed in the hose, which when moved with a finger, it turned out to be a hair".This was observed before patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received for evaluation.A visual inspection was performed, and a loose hair inside the opened primary packaging was observed.The reported condition was verified.The cause of the reported condition could not be determined; however, the possible cause was a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|