Model Number CC-811 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
D10 concomitant products: cc-811, cc-811 caprosyn* 2-0 und 75cm gs21 x36 (lot#:d2f1907y); cc-811, cc-811 caprosyn* 2-0 und 75cm gs21 x36 (lot#:d2f1907y); cc-811, cc-811 caprosyn* 2-0 und 75cm gs21 x36 (lot#:d2f1907y) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a procedure, when removing the suture from the package, the four needles were detached from the thread.
|
|
Manufacturer Narrative
|
H3: evaluation summary: medtronic conducted an investigation based upon all information received.Two devices were available for evaluation.Visual inspection noted two sutures and two needles.One needle was returned disengaged from the suture.The returned suture was noted to be degraded.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.The other needle was returned attached to the suture.It was reported that the needles were detached from the thread.The reported issue was confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: the suture was degraded.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
According to the reporter, during a procedure, when removing the suture from the package, the four needles were detached from the thread.Another device was used to complete the case.There was no patient injury.
|
|
Search Alerts/Recalls
|
|
|