A philips remote service engineer (rse) remotely interviewed the customer who was onsite.The rse explained, dependent on configuration settings for spo2 desaturation/low limit delays, the unit could take a while for the spo2 red level alarm to activate.Also, the latching behavior would dictate how the alarm was displayed.The philips technical consultant (tc) assigned to the case, indicated the customer was unaware of the default alarm delay time and the logs the customer inspected found there was no trouble found.The philips clinical specialist (cs) further assisted the customer with an explanation of the default alarm delay time and there were no further issues reported.The philips cs offered an educator on site to review with staff.The customer declined and stated that it was not necessary.No further investigation or action is warranted at this time.H3 other text : on site work was not required, clinical helped the customer with this matter.
|