Model Number 865240
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/05/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the intellivue mx800 patient monitor indicating spo2 parameter did not alarm for desaturation of 70%.The device was in use on patient at time of event, there was no adverse event reported.
 
Manufacturer Narrative
A philips remote service engineer (rse) remotely interviewed the customer who was onsite.The rse explained, dependent on configuration settings for spo2 desaturation/low limit delays, the unit could take a while for the spo2 red level alarm to activate.Also, the latching behavior would dictate how the alarm was displayed.The philips technical consultant (tc) assigned to the case, indicated the customer was unaware of the default alarm delay time and the logs the customer inspected found there was no trouble found.The philips clinical specialist (cs) further assisted the customer with an explanation of the default alarm delay time and there were no further issues reported.The philips cs offered an educator on site to review with staff.The customer declined and stated that it was not necessary.No further investigation or action is warranted at this time.H3 other text : on site work was not required, clinical helped the customer with this matter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX800 PATIENT MONITOR
Type of Device
INTELLIVUE MX800 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18842124
MDR Text Key336980406
Report Number9610816-2024-00114
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838020733
UDI-Public00884838020733
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865240
Device Catalogue Number865240
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/05/2024
Initial Date FDA Received03/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-