Model Number 471405-06
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument for the failure analysis; however, the investigation is in progress.A supplemental mdr will be submitted if any additional information is received.
 
Event Description
It was reported that during a da vinci-assisted benign hysterectomy surgical procedure, the customer was unable to remove a force bipolar instrument from cannula due to part of instrument catching the distal lip of cannula.The cannula was removed from patient with instrument intact.Pin gauge passed through cannula after removal of instrument with no issue.The procedure was completed with no reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
FORCE BIPOLAR
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18842838
MDR Text Key337488786
Report Number2955842-2024-11963
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874120767
UDI-Public(01)00886874120767(11)230615(10)K12230615(91)0001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471405-06
Device Catalogue Number471405
Device Lot NumberK12230615 0145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/15/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexFemale
-
-