C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1608062 |
Device Problems
Stretched (1601); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a port placement procedure, the guidewire was allegedly stuck in the punch needle.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one powerport clearvue isp implantable port kit was returned for evaluation.Gross visual, microscopic visual, functional and dimensional evaluations were performed on the returned device.An attempt to remove the loaded guidewire from the introducer needle was performed and was successful after resistance was felt and uncoiling was noted throughout the distal end portion of the guidewire.Therefore, the investigation is confirmed for the reported stuck guidewire in needle as even though the loaded guidewire removed from the introducer needle was successful but had a resistance and identified uncoiling issues.Based on the measurements recorded during sample evaluation and applicable drawings, no measurements recorded could be confirmed to be out of tolerance.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.(expiration date: 05/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Search Alerts/Recalls
|
|
|