U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem Bacterial Infection (1735)
Event Date 05/12/2014
Event Type  Death  
Event Description
Complainant alleges through their counsel a mycobacterium chimaera infection acquired post-operatively following a heart valve surgery (date unknown) in which a heater-cooler system 3t device was used.Patient died.
 
Manufacturer Narrative
D.4.Serial number is unknown.This information will be provided in a supplemental report if made available.H.4.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in coventry, united kingdom.Through follow up communication, livanova learned that possible involved device serial numbers could be the following: (b)(6) (model: 16-02-80).No other information has been made available from the customer.Device history records (dhrs) review of possible involved device serial numbers has been completed and did not identify any deviation or non-conformity relevant to the reported issue.No device between the ones possible involved and in use at the hospital at the time of surgery (2014), was equipped with vacuum and sealing kit since upgrade activity started in 2017.No further investigation is possible.Source of patient contamination remains unknown and a direct relationship between the reported adverse event and the device could not be established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18844763
MDR Text Key336963422
Report Number9611109-2024-00115
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/05/2024
Initial Date FDA Received03/06/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
Patient Outcome(s) Death;