• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING Back to Search Results
Model Number 865350
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2024
Event Type  malfunction  
Event Description
The customer reported that the monitor did not alarm properly; the red alarm did not activate only yellow alarm although patient went into ventricular tachycardia (v-tach).The device was in use on patient at time of event, there was no adverse event reported.A philips remote service engineer (rse) spoke with the customer and the customer alleged that the monitor did not alarm properly; the red alarm did not activate only yellow alarm although patient went into v-tach¿.The rse informed that the customer needed configuration password for mx40.The rse gave the customer passwords for mx40's only to resolve the reported issue.A good faith effort (gfe) was made to obtain additional information and resolution for the case.In response, it was noted that the alarm issue did not cause any patient harm and configuration password resolved the reported issue.Based on the information available, the cause of the reported problem was due to the configuration issue of the device.The reported problem was confirmed.The engineer provided the configuration password to the customer to resolve the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18845900
MDR Text Key337695478
Report Number1218950-2024-00163
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082236
UDI-Public00884838082236
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/12/2024
Initial Date FDA Received03/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-