• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VH
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Event Description
It was reported that, during reprocessing, the videoscope tested positive for 3 colony forming units (cfus) of an unspecified microorganism.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Manufacturer Narrative
The suspect device has been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The legal manufacture reviewed the customer provided the cleaning disinfection and sterilization (cds) processes where the following deviation from instructions for use (ifu) was confirmed: the distal end was not flushed with maj-2319.The hygiene microbiological investigation report indicated the channels of the scope were cultured and less than 1 cfu of any microbes were found.The results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.The device evaluation found the connecting tube collapsed, universal cord protector was damaged, and the distal end biopsy channel were kinked.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on a correction to the legal manufacture's (lm) review of the customer completed cleaning disinfection and sterilization (cds) checklist and the lms final investigation.Correction to h10 of the initial report.The reported deviation in reprocessing was inadvertently added.The question answered by the user facility was for tjf model scopes and therefore not applicable as the subject device is a model cyf-vh.The lm reviewed the customer provided the cds processes where no obvious deviations from instructions for use (ifu) were identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.¿cysto-nephro videoscope olympus cyf-vh olympus cyf-vhr reprocessing manual¿ describes the reprocessing procedures for cyf-vh.The reported phenomenon might be prevented by following these descriptions.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18845904
MDR Text Key337764521
Report Number3002808148-2024-02084
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170411250
UDI-Public04953170411250
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K221683
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/11/2024
Initial Date FDA Received03/06/2024
Supplement Dates Manufacturer Received03/21/2024
Supplement Dates FDA Received03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ANIOXYDE 1000
-
-