• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625374160
Device Problems No Device Output (1435); Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/24/2023
Event Type  malfunction  
Manufacturer Narrative
The coaguchek xs meter serial number was (b)(6).Per product labeling, error 4 indicates the test strip is unusable and to test using a new test strip.The meter and strips have been requested for return.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." e3 - occupation was patient/consumer.
 
Event Description
There was an allegation of error 4 and questionable results from the coaguchek xs meter.At 11:00 am, the result from the meter was 3.2 inr.At 1:00 pm, the result from a laboratory using an unknown reagent was 4.2 inr.The therapeutic range was 2.5 inr-3.5 inr.The testing frequency was not provided.
 
Manufacturer Narrative
The reporter's meter was received for investigation.A leakage current was detected on the pcb (printed circuit board) caused by contamination due to improper handling or maintenance.There was no malfunction of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18846944
MDR Text Key337009280
Report Number1823260-2024-00666
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702127104
UDI-Public00365702127104
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04625374160
Device Lot Number71519022
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/09/2024
Initial Date FDA Received03/06/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received05/03/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ASPIRIN.; ATORVASTATIN.; HYDROCODONE.; LISINOPRIL.; METHOCARBAMOL.; WARFARIN.
Patient Age55 YR
Patient Weight88 KG
-
-