Manufacturer's investigation conclusion: the reported events of the centrimag console¿s internal battery taking longer to charge than expected and an alarm that was unable to be silenced were not confirmed.The centrimag console (serial number (b)(6) ) was not returned for analysis, and no log files were associated with the reported event.Per additional information, the alarm was unable to be identified and occurred before the console was in patient use.It was also stated that the console functioned as intended after being powered on again.The root causes of the reported events were unable to be conclusively determined through this analysis.Review of the device history record for centrimag console, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual (rev.L) section 3 "warnings and precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual (rev.L, table 6: symbols on the console, monitor, and flow probes) shows that a flashing or blinking battery light indicates that the battery is charging.The expected charge time for the console¿s internal battery is ~4 hours to ~90% charge, and ~5 hours to 100% charge.The 2nd generation centrimag system operating manual (rev.L) section 11.1-"appendix i ¿ console alarms and alerts" contains a list of console alarms and alerts, as well as appropriate operator response to these events.No further information was provided.The manufacturer is closing the file on this event.
|