THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 201-30300
Device Problems Use of Device Problem (1670); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2024
Event Type  malfunction  
Event Description
It was reported that the customer could see the battery light blinking when the console was continuously charged for a few days.There was also an audible alarm that they were unable to mute.The console was exchanged.The exact type of alarm was unknown, but s1: power on self-test fail was suspected as the device was checked out and was "okay".Motor mfr # 3003306248-2024-00427.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported events of the centrimag console¿s internal battery taking longer to charge than expected and an alarm that was unable to be silenced were not confirmed.The centrimag console (serial number (b)(6) ) was not returned for analysis, and no log files were associated with the reported event.Per additional information, the alarm was unable to be identified and occurred before the console was in patient use.It was also stated that the console functioned as intended after being powered on again.The root causes of the reported events were unable to be conclusively determined through this analysis.Review of the device history record for centrimag console, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual (rev.L) section 3 "warnings and precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual (rev.L, table 6: symbols on the console, monitor, and flow probes) shows that a flashing or blinking battery light indicates that the battery is charging.The expected charge time for the console¿s internal battery is ~4 hours to ~90% charge, and ~5 hours to 100% charge.The 2nd generation centrimag system operating manual (rev.L) section 11.1-"appendix i ¿ console alarms and alerts" contains a list of console alarms and alerts, as well as appropriate operator response to these events.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18849544
MDR Text Key337808971
Report Number3003306248-2024-00426
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K131179
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number201-30300
Device Lot NumberL06358
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/16/2024
Initial Date FDA Received03/06/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/02/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse