It was reported that the patient presented in clinic for a follow-up.Upon interrogation, it was noted that the leadless pacemaker (lp) exhibited undersensing and loss of capture.A chest x-ray was performed and confirmed the lp dislodged to the pulmonary artery.The lp was explanted but and was replaced with a dual chamber intravenous pacemaker three days later.During explant, it was noted the lp helix had stretched.The patient was stable.
|
The reported event of stretched helix was confirmed.The reported event of under sensing, failure to capture, and dislodgement were not confirmed.Final analysis found blood and that the helix was stretched out of specification.The helix elongation is consistent with having occurred during procedure.Further analysis performed found no anomalies contributing to reported event of sensing or capturing problem.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.
|