• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC. EFFICIA DFM100; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC. EFFICIA DFM100; DEFIBRILLATOR Back to Search Results
Model Number 866199
Device Problems Electrical /Electronic Property Problem (1198); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2024
Event Type  malfunction  
Event Description
It was reported the device loudspeaker was damaged.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model# 861290) and will be reported in the united states under device model# 861290.
 
Manufacturer Narrative
This report is based on information provided by philips field service engineer (fse) and has been investigated by the philips complaint handling team.Philips received a complaint on the efficia dfm100 defibrillator indicating that the device has a damaged speaker.The fse evaluated the device on site and determined this was a malfunction of the speaker assembly.The fse replaced the speaker assembly resolving the reported issue.The device remains at the customer site and no further evaluation is warranted at this time.Based on the information available and the testing conducted, the cause of the reported problem was a malfunction of the speaker assembly.The reported problem was confirmed.The fse replaced the speaker assembly resolving the reported issue.It has been concluded that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFICIA DFM100
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH  
Manufacturer Contact
feng she
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key18853001
MDR Text Key337473137
Report Number3030677-2024-00855
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866199
Device Catalogue Number866199
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/08/2024
Initial Date FDA Received03/06/2024
Supplement Dates Manufacturer Received02/08/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-