Model Number VTICM513.2
Device Problems Improper or Incorrect Procedure or Method (2017); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2023
Event Type  malfunction  
Manufacturer Narrative
A4-a6:unk.H6: user error.Claim# (b)(4).
 
Event Description
The reporter indicated a 13.2mm vticm513.2 implantable collamer lens of a -12.5/4.5/090 (sphere/cylinder/axis), tore/broke during loading.On (b)(6) 2024, a replacement lens was implanted into the patient's right eye (od) and the problem was resolved.Cause of the event was a user error, with the device not failing to perform as intended.Status of the eye is reported as, "eye is quiet.".
 
Manufacturer Narrative
Originally the claim was determined to be reportable, but upon further review was determined not mdr reportable due to no patiant contact and cause of event was "user error".(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
800 royal oaks dr. suite: 210
monrovia, CA 91016
6263037902
MDR Report Key18853002
MDR Text Key337087124
Report Number2023826-2024-01070
Device Sequence Number1
Product Code QCB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVTICM513.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/09/2024
Initial Date FDA Received03/06/2024
Supplement Dates Manufacturer Received05/15/2024
Supplement Dates FDA Received05/15/2024
Date Device Manufactured11/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL - SFC-45 - LOT# UNK; FOAM TIP PLUNGER - LOT# UNK; INJECTOR MODEL: MSI-PF - LOT# UNK
Patient Age25 YR
Patient SexFemale
-
-