• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD VERITOR ¿ SARS-COV-2 & FLU A+B; DEVICES DETECTING INFLUENZA A, B, AND C VIRUS ANTIGENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD VERITOR ¿ SARS-COV-2 & FLU A+B; DEVICES DETECTING INFLUENZA A, B, AND C VIRUS ANTIGENS Back to Search Results
Catalog Number 256088
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Event Description
It was reported while using the bd veritor ¿ sars-cov-2 & flu a+b the user received discrepant results for a symptomatic patient.The initial result was covid positive while upon repeat the result was flu a positive.No health impact or consequence reported.
 
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
H.6.Investigation summary: this statement summarizes the investigation results regarding a complaint that alleges discrepant result when using bd veritor¿ sars-cov-2 and flu a+b ass (material#: 256088), batch number 3216617.The customer reported that they were receiving covid positive, flu a negative and flu b negative when testing a symptomatic patient.They then retested the same sample in a different device and received a flu a positive result.Bd quality performs a systematic approach to investigate all complaints.This approach involves review of manufacturing batch history records, visual inspection of retention samples, and visual inspection of customer returned samples, if applicable.Batch history record (bhr) review and retain sample testing were performed on the batch number provided.The results were acceptable, and no relevant issues were found.No return samples were received; therefore, no return sample analysis could be performed.However, a photograph was returned and showed one device with covid positive, flu a negative, flu b negative result and the other device with flu a positive, flu b negative and covid negative results.This complaint was confirmed based on the photographic evidence.The root cause could not be determined.A trend analysis for discrepant result was conducted, no adverse trend was identified.There were no corrective actions taken at this time.
 
Event Description
It was reported while using the bd veritor ¿ sars-cov-2 & flu a+b the user received discrepant results for a symptomatic patient.The initial result was covid positive while upon repeat the result was flu a positive.No health impact or consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VERITOR ¿ SARS-COV-2 & FLU A+B
Type of Device
DEVICES DETECTING INFLUENZA A, B, AND C VIRUS ANTIGENS
Manufacturer (Section D)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH 
Manufacturer (Section G)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH  
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18856050
MDR Text Key337777537
Report Number3006948883-2024-00035
Device Sequence Number1
Product Code QMN
UDI-Device Identifier00382902580881
UDI-Public(01)00382902580881
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number256088
Device Lot Number3216617
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/13/2024
Initial Date FDA Received03/07/2024
Supplement Dates Manufacturer Received06/27/2024
Supplement Dates FDA Received07/23/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-